Keterangan :
3 channel inputs for the simultaneous recording of Leakage Current, Load Current and Voltage
• Power Quality analysis
( only on KEW 5020)
( Power Quality: Reference voltage, Swell, Dip, Short power Interruptions)
Large Capacity for storing 60, 000 data points
60, 000 data points can be recorded when 1ch is used, and when all the three channels are used, 20, 000 data points per channel can be recorded
Lowpass Filter will filter out the harmonics
( Cutoff Frequency = Approx. 160Hz)
LED flickers when the preset current / voltage value is exceeded
CALL: Confirmation of recorded data
• The following can be displayed: number of recorded data points, ( max+ min+ peak) value for each channel complete with time/ date information in the Normal recording mode. ( Detected values ( i.e. when values are outside preset limits ) can be displayed in other recording modes)
• RECALL: The last 10 recorded data points including time/ date can be recalled on the logger display
Selection of One-time mode or Endless mode
• One-time on: Recording will stop when memory is used up
• One-time off: Overwrite the old data, and store the latest data
Non Volatile Memory
Recorded data will be retained even if the batteries are exhausted or replaced due to the presence of a nonvolatile memory ( guaranteed for 10 years)
Battery power indicator
Indicates battery voltage in 4-levels
( It is possible to use the logger for a further approx. 24 hours even after the warming symbol is flashing)
The userfriendly PC software " KEW LOG Soft 2" is supplied
• Supplied with the user friendly software " KEW LOG Soft 2" . This permits editing, analysis and graphical display of data.
• The recorded data is downloadable onto a PC via USB cable.
• Variation of the measured voltage and current data can be confirmed simultaneously on the PC display monitor. ( only on KEW 5020)
• Simplified Power Integration
( The " KEW LOG Soft 2" uses current and voltage recorded to calculate the integral power consumption)
• Continuous measuring time: Approx. 10 days ( Alkaline Battery)
Normal Recording Mode
( AC 50/ 60Hz, Sine wave, Input: 10% or more of the range at CH1)
Range RMS Accuracy
100.0mA | ± 2.0% rdg± 0.9% f.s.+ Accuracy of Sensor
Other ranges | ± 1.5% rdg± 0.7% f.s.+ Accuracy of Sensor
Crest Factor | 2.5 or less: RMS accuracy ( sine) + 2% rdg + 1% f.s.
* Max, Min and Instant Peak values in Normal Recording mode are just reference values:
their accuracies aren' t guaranteed.
Trigger Recording Mode ( AC 50/ 60Hz sine wave)
Range Accuracy
100.0mA | ± 3.5% rdg± 2.2% f.s.+ Accuracy of Sensor
Other ranges | ± 3.0% rdg± 2.0% f.s.+ Accuracy of Sensor
Capture/ Power Quality Analysis Recording Mode
Range Accuracy
100.0mA | ± 3.0% rdg± 1.7% f.s.+ Accuracy of Sensor
Other ranges | ± 2.5% rdg± 1.5% f.s.+ Accuracy of Sensor
KEW 5010 KEW 5020
Recording Mode Normal, Trigger, Capture Normal, Trigger, Capture, Power Quality Analysis
Operating System Successive Approximation
( CH1 single synchronized sampling)
Rated max. working voltage AC 9.9Vrms, 14V peak value
Number of input channel 3ch
Measuring method True RMS
RMS measuring interval approx. 100ms
Sampling interval
Normal/ Trigger mode approx. 1.65ms/ CH
Capture mode approx. 0.55ms ( waveform: at every 1.1ms)
P.Q.A. mode - approx. 0.55ms
Low battery warning Battery mark display ( in 4 levels)
Over-range indication " OL" mark is displayed when exceeding the measuring range
Auto power off Power-off function operates automatically after a switch remains for 3min. ( when recording is stopped)
Location for use Indoor use, Altitude up to 2000m
Operating temperature & humidity range -10º C~ 50º C / Relative humidity 85% or less ( no condensation)
Battery DC6V: Alkaline battery ( LR6) × 4pcs / External supply DC9V ( Special AC Adapter)
Possible measurement time Approx. 10 days ( with alkaline LR6 batteries)
Applicable standards IEC 61010-1: 2001 CAT.III 300V Pollution degree2
IEC 61326 ( EMC standard)
Withstand voltage AC 3540V ( RMS 50/ 60Hz) / for 5 sec.
Dimension 111( L) × 60( W) × 42( D) mm
Weight Approx. 265g
Accessories Alkaline battery LR6× 4pcs, 9118 ( Carrying case)
PC software " KEW LOG Soft 2" , 7148 ( USB cable)
Instruction manual, Quick manual, Install manual,
USB Notice sheet
Optional 8146/ 8147/ 8148 ( Leakage & Load current Clamp Sensor) ,
8121/ 8122/ 8123 ( Load current Clamp Sensor) ,
8129 ( Flexible Clamp Sensor) ,
8309 ( Voltage Sensor: only KEW 5020) , 8320 ( AC Adapter) ,
9135 ( Carrying bag) , 9119 ( Hard carry case) ,
7185 ( Extension cable for sensor)